Rubidium-82 Generator Yield and Efficiency for PET Perfusion Imaging: Comparison of Two Clinical Systems

Authors: Ali Ahmadi, Ran Klein, Howard C Lewin, Rob S.B. Beanlands, Robert A. deKemp

Introduction. Strontium-82/Rubidium-82 (82Sr/82Rb) generators are used widely for positron emission tomography (PET) imaging of myocardial perfusion. In this study, the 82Rb isotope yield and production efficiency of two FDA-approved 82Sr/82Rb generators were compared. Methods. N=515 sequential daily quality assurance (QA) reports from 9 CardioGen-82® and 9 RUBY-FILL® generators were reviewed over a period of 2 years. A series of test elutions was performed at different flow-rates on the RUBY-FILL® system to determine an empirical correction-factor used to convert CardioGen-82® daily QA values of 82Rb activity (dose-calibrator ‘maximum’ of 50 mL elution at 50 mL/min) to RUBY-FILL® equivalent values (integrated ‘total’ of 35 mL elution at 20 mL/min). The generator yield (82Rb) and production efficiency (82Rb yield / 82Sr parent activity) were measured and compared after this conversion to a common scale. Results. At the start of clinical use, the system reported 82Rb activity from daily QA was lower for CardioGen-82® vs RUBY-FILL® (2.3 ± 0.2 vs 3.0 ± 0.2 GBq, p<0.001) despite having similar 82Sr activity. Dose-calibrator ‘maximum’ (CardioGen-82®) values were found to under-estimate the integrated ‘total’ (RUBY-FILL®) activity by ~24% at 50 mL/min. When these data were used to convert the CardioGen-82 values to a common measurement scale (integrated total activity) the CardioGen-82® efficiency remained slightly lower than the RUBY-FILL® system on average (88 ± 4% vs 95 ± 4%, p<0.001). The efficiency of 82Rb production improved for both systems over the respective periods of clinical use. Conclusions. 82Rb generator yield was significantly under-estimated using the CardioGen-82® vs RUBY-FILL® daily QA procedure. When generator yield was expressed as the integrated total activity for both systems, the estimated 82Rb production efficiency of the CardioGen-82® system was ~7% lower than RUBY-FILL® over the full period of clinical use.

Comments: 28 Pages.

Download: PDF

Submission history

[v1] 2020-04-20 14:57:40

Unique-IP document downloads: 233 times

Vixra.org is a pre-print repository rather than a journal. Articles hosted may not yet have been verified by peer-review and should be treated as preliminary. In particular, anything that appears to include financial or legal advice or proposed medical treatments should be treated with due caution. Vixra.org will not be responsible for any consequences of actions that result from any form of use of any documents on this website.

Add your own feedback and questions here:
You are equally welcome to be positive or negative about any paper but please be polite. If you are being critical you must mention at least one specific error, otherwise your comment will be deleted as unhelpful.

comments powered by Disqus